Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1164220180300010027
Journal of Korean Society for Radiotherapeutic Technology
2018 Volume.30 No. 1 p.27 ~ p.34
Effect of Carbon Couch Side Rail and Vac-lok In case of Lung RPO irradiation
Kim Seok-Min

Kwak Geun-Tak
Lee Seung-Hun
Kim Jung-Soo
Kwon Hyoung-Chul
Kim Yang-Soo
Lee Sun-Young
Abstract
Purpose : To evaluate the effect of carbon couch side rail and vacuum immobilization device in case of lung RPO irradiation.

Materials and Methods : The 10, 20, 30 mm thickness of vac-lok¡¯s right side were obtained. To measure of doses, glass dosimeters were used and measured reference point is left lung center at the phantom. A, B, C, and D points are left, right, down, and up directions based on the center point. In the state of Side-Rail-Out, place the without vac-lok, with the thickness of 10, 20, and 30 mm vac-lok. After the glass dosimeters was inserted in center, A, B, C, and D points, 100 MU of 6 MV X-ray were irradiated to the referenced center point in the condition of 10 x 10 cm2 field size, SAD 100 cm, gantry angle 225, 300 MU/min dose rate. Five measurements were made for each point. In the state of Side-Rail-In, five measurement were made for each point under the same conditions. The average is measured on each of the five Side-Rail-Out and Side-Rail-In measurements.

Results : In the presence of side rail, the dose reduction ratio was -11.8 %, -12.3 %, -4.1 %, -12.3 %, -7.3 % for each A, B, C, and D points. In the state of Side-Rail-Out, the dose reduction ratio for the using 10 mm thickness of vac-lok was -0.9 % than without vac-lok. The dose reduction ratio for the using 20 mm thickness of vac-lok was -2.0 %, for the using 30 mm thickness of the vac-lok was -3.0 % than without vac-lok. In the state of Side-Rail-In, the dose reduction ratio for the using 10 mm thickness of vac-lok was -1.0 % than without vac-lok. The dose reduction ratio for the using 20 mm vac-lok was -2.1 %, for the using 30 mm vac-lok was -3.0 % than without vac-lok. Based on the value of no vac-lok dose in the Side-Rail-In state, The dose reduction ratios for the using 10 mm, 20 mm and 30 mm thickness of vac-loks In the Side-Rail-Out that the center point were -12.7 %, -13.7 %, -14.2 % and -12.8 %, -13.8 %, -14.5 % respectively at point A. The dose reduction ratios for the same conditions to the B point were -4.9 %, -6.1 %, -7.1 % and -13.4 %, -14.4 %, -15.5 % respectively at point C. The dose reduction ratios for the same conditions to the D point were -8.4 %, -9.0 %, -10.4 % respectively.

Conclusion : The attenuation was caused by presence of side rails and thickness of vac-lok. Pay attention to these attenuation factors, making it a more effective radiation therapy.
KEYWORD
Vac-lok, Side Rail, Immobilization Device, Glass Dosimeter, Rando Phantom
FullTexts / Linksout information
 
Listed journal information