Journal of Korean Society for Radiotherapeutic Technology
2018 Volume.30 No. 1 p.35 ~ p.40
Assessment of Compensator Thickness in Proton Therapy
Park Yong-Soo

Jang Jun-Yeong
Jo Gwang-Hyeon
Park Yong-Cheol
Choi Byeong-Ki
Abstract
Purpose : The range of force differs from the size of proton energy used in our hospital. The compensator enables to change energy size based on distal thickness which also makes changes in dose rate. Therefore, the purpose of this study is to evaluate the effect of changing the thickness of compensator distal on dose range and beam on time.

Subject and Methodology : Five low energy patients who have received proton therapy were selected as subjects for this study. Beam on was checked for the selected patients during the existing therapy. After then, the thickness of distal of compensator was increased by 2 cm up to 14 cm through proton therapy plan system(TPS) for comparative analysis. For the evaluation of dose range, the value of the target¡¯s conformity index(CI) and the maximum dose of rear side target¡¯s organ at risk(OAR) were compared. Furthermore, to evaluate the effect of therapy time, beam on time was compared by making compensator distal in each thickness.

Result : The result of homogeneity index and conformity index of the increased compensator distal showed the same level in all patients. The comparison results of OAR of target rear side showed 7 cGy at spine cord of abdomen at maximum, 88 cGy at eyeball¡¯s RT lens, 391 cGy at RT lens of nasal cavity 51 cGy at trachea of the mediastinum, and 661 cGy at a small bowl of the pelvis. The comparison results of the beam on time showed a reduction from 126 seconds to 62 seconds for the abdomen, from 105 seconds to 37 seconds for the eyeball, from 187 seconds to 134 seconds for nasal cavity, from 100 seconds to 40 seconds for mediastinum, from 440 seconds to 118 seconds for the pelvis.

Conclusion : The research result showed that as the distal thickness of compensator increased, the size of energy increased. In addition, beam on decreased due to the increase of dose rate. It is expected that the result would help reduce the treatment time and increase the convenience of patients if it is applied to liver patients who need respiratorygated therapy and pediatric patients. However, distal penumbra increased as the size energy increased. Therefore, in treating cases where OAR is in the vicinity of the target rear side, the influence of penumbra should be taken into account in adjusting thickness level of the compensator in proton therapy plan.
KEYWORD
Compensator, Distal thickness, Beam on time, Energy, Dose rate