Journal of Korean Society for Radiotherapeutic Technology
2018 Volume.30 No. 1 p.117 ~ p.128
Usability Assessment of Plan Transfer between TOMO HAD and Radixact : Planning Study
An Ye-Chan

Kim Jong-Sik
Kwon Dong-Yeol
Kim Jin-Man
Choi Byeong-Ki
Abstract
Purpose : To evaluate the usability of plan transfer between TOMO HD and Radixact, we compared the differences of dose in transferred plans by evaluating the dose of normal organ and target. TOMO HDA and Radixact. The completed plans were transferred each other and we compared the differences of dose by evaluating the DVH of each plans.

Materials and Methods : We planned 4 different plans assuming the treatment of 2 cases in Head and Neck Cancer and 2 cases Prostate cancer. Each plan was designed so that 95 % of the prescription dose was irradiated over 99 % of the target volume, and the normal organ constraints dose was based on the SMC tolerance dose protocol. Each plan was transferred to each equipment and DVH(dose volume histogram) analysis of the transferred plans was compared and evaluated.

Results : The Mean dose of CTV and GTV was increased and decreased in the transferred plans, but there was no significant differences. The target coverage of CTV and GTV was decreased in all cases of transferred plans from TOMO HAD to Radixact, and the change of CI and HI in CTV was within 0.1. Normal organ dose was increased in most cases when transferring from HAD to Radixact in both treatment plans.

Conclusion : According to the results of this experiment, the target coverage was above the standard and the normal organ dose was almost same or decreased when transferring the plans from Radixact to HDA equipment. However the target coverage was reduced when transferring the plans from HDA to Radixact and there was an increase in dose in normal organs that could cause sever side effects such as Optic Chiasm (Dmax1.38 Gy), Bladder (Dmax3.07 Gy), Penile Bulb (Dmax1.14 Gy). Therefore, it is necessary to pay attention to the dose change when transferring the plan and one-time transfer due to equipment inspection will be useful for efficient radiation therapy, but if the transferred treatment plans continue for several consecutive days, the treatment plan should be resumed.
KEYWORD
TOMO HDA, Radixact, Transfer, Plan Comparison