Journal of Korean Society for Radiotherapeutic Technology
2018 Volume.30 No. 1 p.191 ~ p.199
Evaluation of a colloid gel(Slime) as a body compensator for radiotherapy
Lee Hun-Hee

Kim Chan-Kyu
Song Kwan-Soo
Bang Mun-Kyun
Kang Dong-Yun
Shin Dong-Ho
Lee Doo-Hyeon
Abstract
Purpose : In this study, we evaluated the usefulness of colloid gel(slime) as a compensator for irregular patient surfaces in radiation therapy.

Materials and Methods : For this study, colloid gel suitable for treatment was made and four experiments were conducted to evaluate the applicability of radiation therapy. Trilogy(Varian) and CT(SOMATOM, Siemens) were used as treatment equipment and CT equipment. First, the homogeneity according to the composition of colloid gel was measured using EBT3 Film(RIT). Second, the Hounsfield Unit(HU) value of colloid gel was measured and confirmed by CRIS phantom, Eclipse RTP(Eclipse 13.1, Varian) and CT. Third, to measure the deformation and degeneration of colloid gel during the treatment period, it was measured 3 times daily for 2 weeks using an ion chamber(PTW-30013, PTW). The fourth experiment was compared the treatment plan and measured dose distributions using bolus, rice, colloid gel and additional, dose profiles in an environment similar to actual treatment using our own acrylic phantom.

Result : First experiment, density of the colloid gel cases 1, 2 and 3 was 1.02 g/cm3, 0.99 g/cm3 and 0.96 g/cm3. When the homogeneity was measured at 6 MV and 9 MeV, case 1 was more homogeneous than the other cases, as 1.55 and 1.98. In the second experiment, the HU values of case 1, 2, 3 were 15 and when the treatment plan was compared with the measured doses, the difference was within 1 % at all 9, 12 MeV and a difference of -1.53 % and -1.56 % within the whole 2 % at 6 MV. In the third experiment, the dose change of colloid gel was measured to be about 1 % for 2 weeks. In the fourth experiment, the dose difference between the treatment plan and EBT3 film was similar for both colloid gel and bolus, rice at 6 MV. But colloid gel showed less dose difference than bolus and rice at 9 MeV. Also, dose profile of colloid gel showed a more uniform dose distribution than the bolus and rice.

Conclusion : In this study, the density of colloid gel prepared for radiation therapy was 1.02 g/cm3 similar to the density of water, and alteration or deformation was not observed during the radiotherapy process. Although we pay attention to the density when manufacturing colloid gel, it is sufficient in that it can deliver the dose uniformly through the compensation of the patient¡¯s body surface more than the bolus and rice, and can be manufactured at low cost. Further studies and studies for clinical applications are expected to be applicable to radiation therapy.
KEYWORD
Colloid gel(Slime), EBT3 Film, Compensator