Journal of Korean Society for Radiotherapeutic Technology
2019 Volume.31 No. 1 p.17 ~ p.24
Assessment on Accuracy of Stereotactic Body Radiation therapy (SBRT) using VERO
Lee Wi-Yong

Kim Hyun-Jin
Yun Na-Ri
Hong Hyo-Ji
Kim Hong-Il
Baek Seung-Wan
Abstract
Purpose : The present study aims to assess the level of coherency and the accuracy of Point dose of the Isocenter of VERO, a linear accelerator developed for the purpose of the Stereotactic Body Radiation Therapy(SBRT).

Materials and Method : The study was conducted randomly with 10 treatment plans among SBRT patients in Kyungpook National University Chilgok Hospital, using VERO, a linear accelerator between June and December, 2018. In order to assess the equipment¡¯s power stability level, we measured the output constancy by using PTWLinaCheck, an output detector. We also attempted to measure the level of accuracy of the equipment¡¯s Laser, kV(Kilo Voltage) imaging System, and MV(Mega Voltage) Beam by using Tofu Phantom(BrainLab, Germany) to assess the accuracy level of geometrical Isocenter. We conducted a comparative analysis to assess the accuracy level of the dose by using an acrylic Phantom(30¡¿30¡¿20 cm), a calibrated ion chamber CC-01(IBA Dosimetry), and an Electrometer(IBA, Dosimetry).

Results : The output uniformity of VERO was calculated to be 0.66 %. As for geometrical Isocenter accuracy, we analyzed the error values of ball Isocenter of inner Phantom, and the results showed a maximum of 0.4 mm, a minimum of 0.0 mm, and an average of 0.28 mm on X-axis, and a maximum of -0.4 mm, a minimum of 0.0 mm, and an average of -0.24 mm on Y-axis. A comparison and evaluation of the treatment plan dose with the actual measured dose resulted in a maximum of 0.97 % and a minimum of 0.08 %.

Conclusion : The equipment¡¯s average output dose was calculated to be 0.66 %, meeting the ¡¾3 % tolerance, which was considered as a much uniform fashion. As for the accuracy assessment of the geometric Isocenter, the results met the recommended criteria of ¡¾1 mm tolerance, affirming a high level of reproducibility of the patient¡¯s posture. The difference between the treatment plan dose and the actual measurement dose was calculated to be 0.52 % on average, significantly less than the 3 % tolerance, confirming that it obtained predicted does. The current study suggested that VERO equipment is suitable for SBRT, and would result in notable therapeutic effect.