Study of the CatcherTM Couch's Usefulness
Um Ki-Cheon

Lee Chung-Hwan
Jeon Soo-Dong
Song Heung-Kwon
Back Geum-Mun
Abstract
Purpose: Recently, A Catcher was added to prevent sagging in Radixact¢ç X9. In this study, We quantitatively compared general couch of Tomo-HDA¢ç with catcher couch of Radixact¢ç X9 using the human phantom and evaluated usefulness of catcher.

Materials and methods: We used rando phantom for phantom study and set the each iso-center of head and neck region and Pelvis region for region parameter. Furthermore, We used hand made low melting point alloys for weight parameter. MVCT(Mega Voltage Computed Tomography) images were acquired for vertical error and rotation(pitch) error measurement increasing weight(A: 15kg, A+B: 30kg, A+B+C: 45kg). We selected 120 patients who has been treated using Tomotherpy machine for patient study. 60 patients has been treated in Tomo-HDA¢ç and the other 60 patients treated in Radixact¢ç X9. In the patient study methods, vertical error and rotation(pitch) error was measured for mean value calculation using MVCT images acquired on first day of radiation therapy.

Result: Result of phantom study, Vertical error and rotation(pitch) error was increased proportionally increased as the weight increases in general couch of Tomo-HDA¢ç. each maximum value was 7.52mm, 0.38¡Æ in head and neck region and 11.94mm, 0.92¡Æ in pelvis region. However, We could confirm that there was stable error range(0.02~0.1mm, 0~0.04¡Æ) in Catcher couch of Radixact¢ç. Result of patient study, The head and neck region was measured 4.79mm 0.33¡Æ lower, and the pelvis region was measured 7.66mm, 0.22¡Æ lower in Catcher couch of Radixact¢ç X9.

Conclusion: In this study, Vertical error and rotation(pitch) error was proportionally increased as the weight increases in general couch of Tomo-HDA¢ç. Especially, The pelvis region error was more increased than the head and neck region error. However, Vertical error and rotation(pitch) error was regularly generated regardless of weight or regions in CatcherTM couch of Radixact¢ç X9 that this study's purpose. In conclusion, CatcherTM couch of Radixact¢ç X9 can minimize mechanical error that couch sagging. Furthermore, The pelvis region is more efficiency than head and neck region. In radiation therapy using Tomotherapy machine, it is regarded that may contribute to minimizing unadjusted pitch error due to characters of Tomotherapy.
KEYWORD
Radixact, Tomotherapy, CatcherTM couch, Set-up
FullTexts / Linksout information
 
Listed journal information