Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1195620080010010010
Clinical and Experimental Otorhinolaryngology
2008 Volume.1 No. 1 p.10 ~ p.14
Clinical Outcomes of Cochlear Reimplantation Due to Device Failure
Kim Chong-Sun

Suh Myung-Whan
Chang Sun-O
Oh Seung-Ha
Kim Dong-Kyu
Abstract
Objectives: The aim of this study was to evaluate the clinical features of cochlear reimplantation due to device failure.

Methods: The medical records of 30 patients who had undergone a revision cochlear implantation were retrospectively reviewed. Causes of revision operations, number of electrode channels inserted, and postoperative speech performances were analyzed.

Results: Device failure (N=12, 38.7%) and hematoma (N=3, 9.6%) were the two most common reasons for revision
surgery. In patients with device failure, the number of electrode channels reinserted was equal to, or more than the number of channels inserted during initial implantation. Speech performance scores remained the same, or improved after reimplantation in patients with device failure.

Conclusion: Device failure was the most common cause of revision operation in patients with cochlear implanttion. Contrary to expectation, new electrodes were fully inserted without difficulty in all reimplantation cases.
Intracochlear damage due to reimplantation appeared to be clinically insignificant.
KEYWORD
Cochlear implantation, Revision, Device failure, Deaf, Reimplantation
FullTexts / Linksout information
 
Listed journal information
SCI(E) ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø