Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1207720100020020125
Clinics in Orthopedic Surgery
2010 Volume.2 No. 2 p.125 ~ p.127
Separation of the Tip of a Coblation Wand within the Knee Joint: A Complication of Arthroscopic Adhesiolysis
Kim Yeub

Kim Ha-Kyung
Yoon Jung-Ro
Muzaffar Nasir
Kim Taik-Sun
Shin Young-Su
Abstract
Coblation devices are now widely used in arthroscopic surgery and they show a very low incidence of intraoperative complications. We experienced a case where the tip of the wand separated and migrated into the posterior knee compartment in an arthrofibrotic knee. The free wand tip was identified and then extricated from the popliteal hiatus of the knee with using C-arm fluoroscopic control. To the best of our knowledge, this is the first report of its kind involving coblation wands. We describe this complication to show that the use of coblation devices can lead to unexpected problems and it is imperative to inspect all instruments before and after each surgical use.
KEYWORD
Arthroscopy, Complication, Coblation wand
FullTexts / Linksout information
 
Listed journal information
MEDLINE ÇмúÁøÈïÀç´Ü(KCI) KoreaMed