Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1207720230150040690
Clinics in Orthopedic Surgery
2023 Volume.15 No. 4 p.690 ~ p.694
Kinematically Aligned Oxford Unicompartmental Knee Arthroplasty Using the Microplasty Instrumentation System
Takafumi Hiranaka

Takaaki Fujishiro
Motoki Koide
Koji Okamoto
Abstract
This technical note demonstrates kinematically aligned Oxford unicompartmental knee arthroplasty using the Microplasty instrumentation system with custom-made devices. The medial joint line is evaluated preoperatively; if it is aligned and parallel with the lateral joint line, they are considered to comprise the coronal knee joint line (CJL). In this case, the coronal inclination of a spoon gauge inserted into the medial joint space indicates the CJL. Otherwise, an accessory spoon is inserted and connected to the medial spoon to refer to the posterior condylar line, which is considered the CJL. The tibial cutting block is then connected without changing the inclination of the spoon and the coronal tilt of the tibial extramedullary rod is adjusted, which is implemented with a custom-made ankle yoke. The remainder of the steps is then identical to the conventional Microplasty procedure. This technique can imitate the cutting line to the CJL, which might be considered ideal from mechanical and kinematic perspectives.
KEYWORD
Knee, Arthroplasty, Oxford, Kinematic, Alignment, Joint line
FullTexts / Linksout information
 
Listed journal information