KMID : 1207720230150050760
Clinics in Orthopedic Surgery
2023 Volume.15 No. 5 p.760 ~ p.769
Accuracy of Patient-Specific Instrument for Cylindrical Axis Implementation in Kinematically Aligned Total Knee Arthroplasty
Kim Kwang-Kyoun

Song Jae-Hwang
Abstract
Background : In kinematically aligned total knee arthroplasty (KA-TKA), the cylindrical axis (CA) is very important in restoring the native joint line and kinematics of the pre-arthritic knee. This study aimed to determine the accuracy of patient-specific instrument (PSI) for restoring the CA for femoral bone resection in KA-TKA.

Methods : Thirty KA-TKAs were performed using a computed tomography (CT)-based PSI system. Data from preoperative CT were reconstructed into three-dimensional (3D) models using 3D-planning software. The CA was created by connecting the centers of each virtual sphere to the medial and lateral femoral condyles using computer software. Femoral bone resection of the distal and posterior condyles was performed parallel to the sagittal planes of the CA. The thickness of the CA-referenced bone resection was determined based on the thickness necessary for the respective regions of the femoral component. The PSI was manufactured to locate the guide pin for a conventional cutting block. The accuracy of PSI for KA-TKA was evaluated as the absolute error between the preoperatively predicted thickness and the intraoperative measurements in each of the four regions, as well as the difference in error between distal-medial (DM) and posterior-medial (PM) and between distal-lateral (DL) and posterior-lateral (PL).

Results : The differences in thickness of bone cut in the DM, DL, PM, and PL were 0.79 ¡¾ 0.39 mm (range, ?1.20 to 1.50), 0.70 ¡¾ 0.42 mm (range, ?1.50 to 1.50), 0.80 ¡¾ 0.46 mm (range, ?0.80 to 1.50), and 0.75 ¡¾ 0.47 mm (range, ?2.10 to 1.40), respectively. There was no significant difference in the thickness error between DM and PM (p = 0.959) and between DL and PL (p = 0.812).

Conclusions : In KA-TKA, PSI was effective for accurate femoral bone resection based on virtually planned thickness.
KEYWORD
Total knee arthroplasty, Patient-specific instrument, Computer-assisted surgery, Kinematic alignment