KMID : 1211620180130010099
Journal of the Korean Society of Physical Medicine
2018 Volume.13 No. 1 p.99 ~ p.105
Concurrent Validity and Clinical Usefulness of Universal Plastic Goniometer for Hip Internal and External Rotation Range Measurement
Kim Yong-Wook

Abstract
PURPOSE: The aim of this study was to evaluate the concurrent validity and clinical usefulness of the universal plastic goniometer to measure the range of motion of the internal and external rotation of the hip joint using the three dimensional motion analysis which can analyze the joints and segment movements in the most objective and quantitative method.

METHODS: Clinical and kinematic data were collected from thirty individuals using a universal plastic goniometer and a ten camera motion analysis system. Passive hip rotation range was obtained three trials for left and right hip joints using two measure methods simultaneously.

RESULTS: There were significant differences between all matching measures of the two measures of internal and external rotation of the hip joint (p<.05). The relationship between the two tests for all measurements of the internal and external rotation of the hip was statistically significant with correlation coefficient form r=.87 to .96. (p<.01).

CONCLUSION: Clinical measurement of the internal and external rotation of the hip using a universal plastic goniometer is effective to assess the hip condition. However, application of universal plastic goniometer requires careful attention in more accurate evaluation and research verification of the internal and external rotation of hip joint.
KEYWORD
3D motion analysis, Goniometry, Hip joint, Range of motion