A Case of Migration of Pipeline Embolization Device Causing Rupture during Treatment of an Unruptured Vertebral Artery Dissecting Aneurysm
Kim Sung-Ho

Yeo Dong-Kyu
Lee Gwang-Soo
Abstract
Endoluminal reconstruction of an intracranial aneurysm using flow-diverting devices, such as the pipeline embolization device (PED), is a new treatment modality with good clinical outcomes. The device was originally indicated for challenging cases, such as wide-necked large or giant aneurysms, and is gaining popularity as a reliable treatment for nearly all intracranial aneurysms. The overall complication rate of flow-diverting devices use is 17.0%, including occlusion of side-branching or perforating arteries, rerupture of the aneurysm, in-stent thrombosis, and, rarely, stent migration. We report a rare complication of the PED: delayed migration of the PED after successful stent implantation during treatment of an unruptured vertebral artery dissecting aneurysm, which resulted in rupture of the aneurysm. Further, we discuss technical steps that can be taken to prevent this potential complication.
KEYWORD
Flow diverting device, Migration, Aneurysm rupture, Complication, Case report