Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0043319990220030294
Archives of Pharmacal Research
1999 Volume.22 No. 3 p.294 ~ p.299
Improved Fluorescent Determination Method of Cellular Sphingoid Bases in High-performance Liquid Chromatography
Yoon Hong-Tak

Yoo Hwan-Soo
Shin Bum-Kyu
Lee Woo-Jin
Kim Hwan-Mook
Hong Seon-Pyo
Moon Dong-Cheul
Lee Yong-Moon
Abstract
Precolumn orthophthaldehyde (OPA) labeling method of sphingoid bases, sphingosine and sphinganine, was investigated to obtain high fluorescent detectability. In order to improve the fluorescent yield, we investigated the optimal solubility of sphingoid bases for five pre-incubation solvents by incorporating the heating procedure before OPA derivatization. The pre-incubation in ethanol prominently increased the fluorescent peak height of OPA derivative for each sphingoid bases in high performance liquid chromatography. About tenfold increase of detectability was archived by pre-incubating lipid extracts pellets in ethanol at for 30 min. Optimal derivatization was performed in 30 min at ambient temperature and the fluorescent intensity of OPA derivative was stable for two weeks at . The detection limit of sphingosine was 0.1 pmol as injected amount. This method was applied to the determination of cellular sphingosine and sphinganine in various human lung cancer cells. This OPA procedure was prospective to be useful for quantitating the amount of sphingoid bases in other cancer cells.
KEYWORD
Sphingosine, Sphinganine, Pe-incubation, OPA, Fluorescent derivatization, HPLC, Lung cancer cells
FullTexts / Linksout information
Listed journal information
SCI(E) MEDLINE ÇмúÁøÈïÀç´Ü(KCI)