3D Stacked Radiation Collimator
|
|
Yoon Dok-Un
Lee Tae-Woong Lee Won-Ho
|
|
Abstract
|
|
|
Multileaf collimators whose Pb leaves are moving in two-dimensional directions have been used. We propose a different concept three-dimensional (3D) collimator with 3D shape that is automatically changeable to modulate the radiation dose even for complex tumors in real time. A voxel collimator, including a hinged Pb plane and a 3D assembly of many voxel collimators, was used. In each frame rotation axis, a motor, which was controlled by a circuit with field-programmable gate array (FPGA) board connected with computer, was operated according to a predetermined plan. Simulations of that, which are generally used for planning, were performed and compared with experimental results.
|
|
KEYWORD
|
|
3D assembly of many voxel collimators, three-dimensional collimatorFPGA
|
|