KMID : 0357520160390010013
|
|
Journal of Radiological Science and Technology 2016 Volume.39 No. 1 p.13 ~ p.17
|
|
Set Up and Operation for Medical Radiation Exposure Quality Control System of Health Promotion Center
|
|
Kim Jung-Su
Jung Hae-Kyoung Kim Jung-Min
|
|
Abstract
|
|
|
In this study, standard model of medical radiation dosage quality control system will be suggested and the useful of this system in clinical field will be reviewed. Radiation dosage information of modalities are gathered from digital imaging and communications in medicine(DICOM) standard data(such as DICOM dose SR and DICOM header) and stored in database.
One CT scan, two digital radiography modalities and two mammography modalities in one health promotion center in Seoul are used to derive clinical data for one month. After 1 months research with 703 CT scans, the study shows CT 357.9 mGy¡¤cm in abdomen and pelvic CT, 572.4 mGy¡¤cm in brain without CT, 55.9 mGy¡¤cm in calcium score/heart CT, screening CT at 54 mGy¡¤cm in chest screening CT(low dose screening CT scan), 284.99 mGy¡¤cm in C-spine CT and 341.85 mGy¡¤cm in L-spine CT as health promotion center reference level of each exam. And with 1955 digital radiography cases, it shows 274.0 mGy¡¤cm2 and for mammography 6.09 mGy is shown based on 536 cases. The use of medical radiation shall comply with the principles of justification and optimization. This quality management of medical radiation exposure must be performed in order to follow the principle. And the procedure to reduce the radiation exposure of patients and staff can be achieved through this. The results of this study can be applied as a useful tool to perform the quality control of medical radiation exposure
|
|
KEYWORD
|
|
Radiation exposure, Patient dose, Quality control, Health promotion, Diagnostic reference levels
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|