KMID : 0358920130400040268
|
|
Journal of the Korean Academy of Pedodontics 2013 Volume.40 No. 4 p.268 ~ p.273
|
|
The Effectiveness of Clinical Sterilization Methods in Dental Air/water Syringes
|
|
Shin Se-young
Yang Yeon-Mi Kim Mi-Ah Kim Jae-Gon Baik Byeong-Ju
|
|
Abstract
|
|
|
This study aims to investigate contamination level and effective clinical disinfection methods of dental air/water syringes (AWS) by using microbial incubation. This study used AWS of ten dental unit chairs of Hospital. Total 180 samples were obtained. There are six groups of samples: non-sterilized (group 1), sanitized with wet-gauze (group 2), sanitized with 78% ethanol sponge for 10 seconds (group 3), sanitized with 78% ethanol sponge for 20 seconds (group 4), sanitized with 1 : 100 diluted High Level Disinfectant (group 5), autoclaved (group 6). Group 1 and 2 showed statistically significant level of CFUs than Group 4, 5 and 6 (p < 0.05). Group 4, 5 and 6 did not show any noticeable CFU. Sanitizing AWS tips with ethanol and High level Disinfectant proves to be a useful and practical method for preventing cross-infections.
|
|
KEYWORD
|
|
Dental air/water syringe, Sterilization methods, Microbial contamination
|
|
FullTexts / Linksout information
|
|
|
|