Journal of the Korean Society for Therapeutic Radiology and Oncology
1988 Volume.6 No. 1 p.101 ~ p.108
Characteristics of 10 MV X-ray Beam from a Mevatron KD Linear Accelerator
Yi Byong-Yong

Lee Myung-Za
Abstract
The beam characteristics and dosimetric measurements of the 10 MV X-ray beam from a Mevatron KD linear accelerator are examined. The Percent Depth Dose (PDD) table and the Tissue Maximum Ratio (TMR) table are taken from measurement as a function of the field size and the depth. The calculated TMR table from PDD table is compared with those from measurement. Other beam characteristics such as output factor, beam profile (including flatness, symmetry and penumbra), wedge, and the variation of Dmax are presented.
KEYWORD
10 MV X-ray, TMR, PDD