KMID : 0364920110360010023
|
|
Journal of Radiation Protection and Research 2011 Volume.36 No. 1 p.23 ~ p.27
|
|
MOSFET Dosimetry for Evaluation of Gonad Shielding during Radiotherapy
|
|
Kim Hwi-Young
Choi Yun-Seok Park So-Yeon Park Yang-Kyun Ye Sung-Joon
|
|
Abstract
|
|
|
In order to confirm feasibility of MOSFET modality in use of in?vivo dosimetry, evaluation of gonad shielding in order to minimize gonadal dose of patients undergoing radiotherapy by using MOSFET modality was performed. Gonadal dose of patients undergoing radiotherapy for rectal cancer in the department of radiation oncology of Seoul National University Hospital since 2009 was measured. 6 MV and 15 MV photon beams emitted from Varian 21EX LINAC were used for radiotherapy. In order to minimize exposed dose caused by the scattered ray not only from collimator of LINAC but also from treatment region inside radiation field, we used box?shaped lead shielding material. The shielding material was made of the lead block and consists of 7.5 cm ¡¿ 9.5 cm ¡¿ 5.5 cm sized case and 9 cm ¡¿ 9.5 cm ¡¿ 1 cm sized cover. Dosimetry for evaluation of gonad shielding was done with MOSFET modality. By protecting with gonad shielding material, average gonadal dose of patients was decreased by 23.07% compared with reference dose outside of the shielding material. Average delivered gonadal dose inside the shielding material was 0.01 Gy. By the result of MOSFET dosimetry, we verified that gonadal dose was decreased by using gonad shielding material. In compare with TLD dosimetry, we could measure the exposed dose easily and precisely with MOSFET modality.
|
|
KEYWORD
|
|
Gonad shielding, MOSFET dosimetry, In-vivo dosimetry
|
|