Dosimetry by Using EBT2 Film for Total Skin Electron Beam Therapy (TSET)
Hwang Ui-Jung

Rah Jeong-Eun
Jeong Ho-Jin
Ahn Sung-Hwan
Kim Dong-Wook
Lee Sang-Yeob
Lim Young-Gyung
Yoon Myong-Geun
Shin Dong-Ho
Lee Se-Byeong
Park Sung-Yong
Pyo Hong-Ryull
Chung Weon-Kuu
Abstract
For treatment of Total Skin Electron beam Therapy (TSET), measurement of dose at various conditions is need on the contrary to usual radiotherapy. When treating TSET with modified Stanford technique based on linear accelerator, the energy of treatment electron beam, the spatial dose distribution and the actual doses deposited on the surface of the patient were measured by using EBT2. The measured energy of the electron beam was agreed with the value that measured by ionization chamber, and the spatial dose distribution at the patient position and the doses at several point on the patient¡¯s skin could be easily measured by EBT2 film. The dose on the patient that was measured by EBT2 film showed good agreement with the data measured simultaneously by TLD. With the results of this study, it was proven that the EBT2 film can be one of the useful dosimeter for TSET.
KEYWORD
TSET, EBT2 film, TLD, Dosimeter, In-vivo