Korean Journal of Radiology 2020 Volume.21 No. 9 p.1036 ~ p.1044
|
|
Feasibility Study of Synthetic Diffusion-Weighted MRI in Patients with Breast Cancer in Comparison with Conventional Diffusion-Weighted MRI
|
|
Choi Bo-Hwa
Baek Hye-Jin Ha Ji-Young Ryu Kyeong-Hwa Moon Jin-Il Park Sung-Eun Bae Kyung-Soo Jeon Kyung-Nyeo Jung Eun-Jung
|
|
Abstract
|
|
|
Objective: To investigate the clinical feasibility of synthetic diffusion-weighted imaging (sDWI) at different b-values in patients with breast cancer by assessing the diagnostic image quality and the quantitative measurements compared with conventional diffusion-weighted imaging (cDWI).
Materials and Methods: Fifty patients with breast cancer were assessed using cDWI at b-values of 800 and 1500 s/mm2 (cDWI800 and cDWI1500) and sDWI at b-values of 1000 and 1500 s/mm2 (sDWI1000 and sDWI1500). Qualitative analysis (normal glandular tissue suppression, overall image quality, and lesion conspicuity) was performed using a 4-point Likert-scale for all DWI sets and the cancer detection rate (CDR) was calculated. We also evaluated cancer-to-parenchyma contrast ratios for each DWI set in 45 patients with the lesion identified on any of the DWI sets. Statistical comparisons were performed using Friedman test, one-way analysis of variance, and Cochran's Q test.
Results: All parameters of qualitative analysis, cancer-to-parenchyma contrast ratios, and CDR increased with increasing b-values, regardless of the type of imaging (synthetic or conventional) (p < 0.001). Additionally, sDWI1500 provided better lesion conspicuity than cDWI1500 (3.52 ¡¾ 0.92 vs. 3.39 ¡¾ 0.90, p < 0.05). Although cDWI1500 showed better normal glandular tissue suppression and overall image quality than sDWI1500 (3.66 ¡¾ 0.78 and 3.73 ¡¾ 0.62 vs. 3.32 ¡¾ 0.90 and 3.35 ¡¾ 0.81, respectively; p < 0.05), there was no significant difference in their CDR (90.0%). Cancer-to-parenchyma contrast ratios were greater in sDWI1500 than in cDWI1500 (0.63 ¡¾ 0.17 vs. 0.55 ¡¾ 0.18, p < 0.001).
Conclusion: sDWI1500 can be feasible for evaluating breast cancers in clinical practice. It provides higher tumor conspicuity, better cancer-to-parenchyma contrast ratio, and comparable CDR when compared with cDWI1500.
|
|
KEYWORD
|
|
Breast cancer, b-value, Diffusion-weighted imaging, MRI, Synthetic MRI
|
|