Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0893319970120030075
Journal of Environmental Toxicology
1997 Volume.12 No. 3 p.75 ~ p.84
Chromatographic Separation of Some Phenol Derivatives Using -Cyclodextrin in Mobile Phase
Moon Young-Ja

Kim Bong-Hee
Abstract
Chromatographic retention behavior and separation of various phenol derivatives on a Partisil 10 ODS 3 column-with mobile phase containing $\alpha$-cyclodextrin-were systematically studied. The decrease in k¡¯ values caused by the addition of cyclodextrins in the mobile phase was based on the formation of an inclusion complex, resulting in weakening of the hydrophobic interaction between solutes and the stationary phase. The content of the organic solvent in the mobile phase also influenced k¡¯ values of the solutes, and k¡¯ values increased with a decrease of the content of organic solvent in the mobile phase. A simple equation has been derived that reveals the hyperbolic dependence of the capacity factor on the total concentration of cyclodextrin. A plot of the reciprocal of the capacity factor against (CD)$_T$ gives a straight line and the dissociation constant, K$_D$, of the inclusion complex can be calculated from the slope. The capacity factor decreased with increasing temperature. The enthalpy was calculated from the slope of van¡¯t Hoff plots. Under optimum conditions, some mixtures of phenol derivatives were able to separated successfully.
KEYWORD
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)