Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1001920150580060513
Journal of Korean Neurosurgical Society
2015 Volume.58 No. 6 p.513 ~ p.517
Endoscopic Fluorescence Angiography with Indocyanine Green : A Preclinical Study in the Swine
Cho Won-Sang

Kim Jeong-Eun
Kim Sae-Hoon
Kim Hee-Chan
Kang Uk
Lee Dae-Sic
Abstract
Objective: Microscopic indocyanine green (ICG) angiography is useful for identifying the completeness of aneurysm clipping and the preservation of parent arteries and small perforators. Neuroendoscopy is helpful for visualizing structures beyond the straight line of the microscopic view. We evaluated our prototype of endoscopic ICG fluorescence angiography in swine, which we developed in order to combine the merits of microscopic ICG angiography and endoscopy.

Methods: Our endoscopic ICG system consists of a camera, a light source, a display and software. This system can simultaneously display real-time visible and near infrared fluorescence imaging on the same monitor. A commercially available endoscope was used, which was 4 mm in diameter and had an angle of 30¡Æ. A male crossbred swine was used.

Results: Under general anesthesia, a small craniotomy was performed and the brain surface of the swine was exposed. ICG was injected via the ear vein with a bolus dose of 0.3 mg/kg. Visible and ICG fluorescence images of cortical vessels were simultaneously observed on the display monitor at high resolution. The real-time merging of the visible and fluorescent images corresponded well.

Conclusion: Simultaneous visible color and ICG fluorescent imaging of the cortical vessels in the swine brain was satisfactory. Technical improvement and clinical implication are expected.
KEYWORD
Indocyanine green fluorescence angiography, Aneurysm surgery, Endoscopy