Zoom-in X-ray Micro Tomography System
|
|
Chun IK
Lee SC/Park JJ/Cho MH/Lee SY
|
|
Abstract
|
|
|
We introduce an x-ray micro tomography system capable of high resolution imaging of a local region inside a small animal. By combining two kinds of projection data, one from a full field-of-view (FOV) scan of the whole body and the other from a limited FOV scan of the region of interest, we have obtained zoomed-in images of the region of interest without any contrast a nomalies. We have integrated a micro tomography system using a micro-focus x-ray source, a $1248\times1248$ flat-panel x-ray detector, and a precision scan mechanism. Using the cross-sectional images taken with the zoom-in micro tomography system, we measured trabecular thicknesses of femur bones in postmortem rats. To compensate the limited spatial resolution in the zoom-in micro tomography images, we used the fuzzy distance transform for the calculation of the trabecular thickness. To validate the trabecular thickness measurement with the zoom-in micro tomography images, we compared the measurement results with the ones obtained from the conventional micro tomography images of the extracted bone samples.
|
|