Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1142720230260020037
Annals of Clinical Microbiology
2023 Volume.26 No. 2 p.37 ~ p.40
False Positive Cases in Automated Blood Culture Systems Due to Hyperleukocytosis

Abstract
Automated blood culture systems are widely used in clinical microbiology laboratories to minimize the workload of laboratory personnel and permit fast turnaround times. However, sometimes false positive signals occur due to leukocytosis, presence of fastidious bacteria, or unexplained causes. We experienced false positive signal in a patient with hyperleukocytosis for the first time since the automated blood culture system was introduced in our hospital
over 20 years ago. We present two case reports with literature review and describe the procedure for dealing with false-positive cases in our hospital.
KEYWORD
Automated blood culture systems, False positive, Hyperleukocytosis
FullTexts / Linksout information
Listed journal information