KMID : 1143620130170020067
Korean Journal of Nuclear Medicine Technology
2013 Volume.17 No. 2 p.67 ~ p.71
A Study on the Use of Active Protocol Using the Change of Pitch and Rotation Time in PET/CT
Jang Eui-Sun

Kwak In-Suk
Park Sun-Myung
Choi Choon-Ki
Lee Hyuk
Kim Soo-Young
Choi Sung-Ook
Abstract
Purpose: The Change of CT exposure condition have a effect on image quality and patient exposure dose. In this study, we evaluated effect CT image quality and SUV when CT parameters (Pitch, Rotation time) were changed.

Materials and Methods: Discovery Ste (GE, USA) was used as a PET/CT scanner. Using GE QA Phantom and AAPM CT Performance Phantom for evaluate Noise of CT image. Images are acquired by using 24 combinations that four stages pitch (0.562, 0.938, 1.375, 1.75:1) and six stages X-ray tube rotation time (0.5s-1.0s). PET images are acquired using 1994 NEMA PET Phantom (18F-FDG 5.3 kBq/mL, 2.5 min/frame). For noise test, noise are evaluated by standard deviation of each image's CT numbers. And then we used expectation noise according to change of DLP (Dose Length Product) to experimental noise ratio for index of effectiveness. For spatial resolution test, we confirmed that it is possible to identify to 1.0 mm size of the holes at the AAPM CT Performance Phantom. Finally we evaluated each 24 image's SUV.

Results: Noise efficiency were 1.00, 1.03, 1.01, 0.96 and 1.00, 1.04, 1.02, 0.97 when pitch changes at the QA Phantom and AAPM Phantom. In case of X-ray tube rotation time changes, 0.99, 1.02, 1.00, 1.00, 0.99, 0.99 and 1.01, 1.01, 0.99, 1.01, 1.01, 1.01 at the QA Phantom and AAPM Phantom. We could identify 1.0 mm size of the holes all 24 images. Also, there were no significant change of SUV and all image's average SUV were 1.1.

Conclusion: 1.75:1 pitch is the most effective value at the CT image evaluation according to pitch change and It doesn't affect to the spatial resolution and SUV. However, the change of rotation time doesn't affect anything. So, we recommend to use the effective pitch like 1.75:1 and adequate X-ray tube rotation time according to patient size.
KEYWORD
Pitch, X-ray tube rotation time, SUV, Noise, Resolution