KMID : 1143620220260010038
Korean Journal of Nuclear Medicine Technology
2022 Volume.26 No. 1 p.38 ~ p.41
The Validation of a Commercial Testosterone RIA Test Kits
Ryu Hwa-Jin

Shin Seon-Yeong
Cho Seong-Uk
Abstract
Purpose: Testosterone is a steroid hormone synthesized by the Leydig cells of the testes in men, and by the adrenal cortex and ovaries in women. Testosterone production is regulated by luteinzing hormone secreted by the anterior pituitary gland. In this experiment, the effectiveness of testosterone radioimmunoassay (RIA) kits produced by three companies was evaluated and compared in case the production of testosterone kits was stopped or supply problems occurred.

Materials and Methods: In October 2021, samples were collected from the patients (n=49) who requested the testosterone RIA test. The experiment was conducted by dividing the patient's sample into low concentration (1.0 ng/mL or less), medium concentration (2.0-4.0 ng/mL) and high concentration (6.0 ng/mL or more). The Testosterone RIA test compared and evaluated the validity of Company A kits used in this hospital and those of Company B and C used in other hospitals. The precision, sensitivity, recovery, linearity and correlation were evaluated for each kit. The testosterone RIA test was carried out in accordance with the insert kit manual for each manufacturer.

Results: As a result of measuring the precision of the intra assay, the Coefficient of Variation (CV) value of the company A kit was high at 11.4% only in the low concentration sample, and in the case of the company B and C kits, the CV value was less than 10% at low, medium, and high concentrations. In the inter-assay precision measurement, the CV value was less than 15% in both A and C kits, but in the case of the B kit, the CV value exceeded 15% at low and medium concentrations. Sensitivity was 0.13 ng/mL for company A, 0.01 ng/mL for company B, and 0.01 ng/mL for company C, and the linearity of all three kits showed excellent linearity. In the case of recovery rate, all of the A, B, and C company kits showed results that were out of 90-110%. In the case of correlation test, when compared with the company A kit currently use in here, the correlation coefficient (R2) value for the company B kit was 0.9508, and for the company C kit was 0.9352

Conclusion: As a result, there was a slight difference in precision at the low concentration sample. The correlation test showed an excellent correlation coefficient. However, it was difficult to secure samples of various concentrations because there were not many tests of testosterone requested at this hospital. So, additional experiments should be carried out by acquiring samples of various concentrations on each laboratory later.
KEYWORD
Testosterone, Evaluation of efficacy, RIA(Radioimmunoassay)