Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 1155820120070020085
Neurointervention
2012 Volume.7 No. 2 p.85 ~ p.92
Endovascular Embolization of Intracranial Aneurysms Using Bare Platinum Axium¢â Detachable Coils: Immediate and Short-Term Follow-up Results from a Multicenter Registry
Kim Byung-Moon

Kim Dong-Joon
Kim Dong-Ik
Jeon Pyoung
Yoon Pyung-Ho
Lee Byung-Hee
Lee Myeong-Sub
Lee Tae-Hong
Byun Jun-Soo
Abstract
Purpose: Axium¢â coils were developed to improve the durability of coil-embolized cerebral aneurysms by increasing packing density. The purpose of this prospective multicenter registry was to evaluate the safety and durability of Axium¢â coils.

Materials and Methods: One hundred twenty-six patients with 135 aneurysms of ¡Â 15 mm in size underwent coil embolization using bare platinum coils, with Axium¢â coils constituting over 50% of the total coil length. Immediate and short-term follow-up results were prospectively registered and retrospectively evaluated.

Results: Of the 135 aneurysms (83 unruptured and 52 ruptured), immediate post-embolization angiography revealed complete occlusion in 80 aneurysms (59.3%), neck remnants in 47 (34.8%), and incomplete occlusion in 8 (5.9%). The mean packing density was 42.8% (range, 9.5 - 90%) with Axium¢â coil length constituting a mean of 87.9% of total coil length. The rate of procedure-related complications was 16.3%. Procedure-related permanent morbidity and mortality rates were 3.2% and 0.8%, respectively. Follow-up catheter or MR angiography, which was available in 101 aneurysms at 6 - 15 months (mean, 7.7 months), revealed stable or improved occlusion in 95 aneurysms and worsening in 6 aneurysms (5.9%). Lower packing density (< 30%) remained the only predictor for anatomical worsening on multivariable logistic regression analysis (P < 0.05).

Conclusion: In this registry, Axium¢â coils showed a relatively low rate of anatomical worsening on short-term follow-up imaging with an acceptable periprocedural safety profile compared to reports of other platinum coils. These results may warrant further study of long-term durability with Axium¢â coils in larger populations.
KEYWORD
Aneurysm, Coiling, Endovascular treatment
FullTexts / Linksout information
Listed journal information
KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø