Journal of Korean Society for Radiotherapeutic Technology
2012 Volume.24 No. 2 p.189 ~ p.196
Evaluation of the Efficiency of the Foxtail Millet Vacuum Cushion in Skin Cancer Radiation Treatment
Choi Shin-Cheol

Lee Kyung-Jae
Jung Sung-Min
Oh Tae-Sung
Park Chong-Yil
Shin Hyun-Kyoh
Abstract
Purpose: The sufficiency of skin dose and the reemergence of patient set-up position to the success of skin cancer radiation treatment is a very important element. But the conventional methods to increase the skin dose were used to vacuum cushion, bolus and water tank have several weak points. For this reason, we producted Foxtail Millet Vacuum Cushion and evaluated the efficiency of the Foxtail Millet Vacuum Cushion in skin cancer
Radiation treatment.

Materials and Methods: We measured absolute dose for 3 materials (Foxtail Millet Vacuum Cushion, bolus and solid water phantom) and compared each dose distribution. We irradiated 6 MV 100 MU photon radiation to every material of 1 cm, 2 cm, 3 cm thickness at three times. We measured absolute dose and compared dose distribution. Finally we inspected the CT simulation and radiation therapy planing using the Foxtail Millet Vacuum Cushion.

Results: Absolute dose of Foxtail Millet Vacuum Cushion was similar to absolute dose of bolus and solid water phantom¡¯s result in each thickness. it Showed only the difference of 0.1~0.2% between each material. Also the same result in dose distribution comparison. About 97% of the dose distribution was within the margin of error in the prescribed ranges (100¡¾R3%), and achieved the enough skin dose (Gross Tumor Volume dose : 100¡¾?5%) in radiation therapy planing.

Conclusion: We evaluated important fact that Foxtail Millet Vacuum Cushion is no shortage of time to replace the soft tissue equivalent material and normal vacuum cushion at the low energy radiation transmittance. Foxtail Millet Vacuum Cushion can simultaneously achieve the enough skin dose in radiation therapy planing with maintaining normal vacuum cushion¡¯ function. Therefore as above We think that Foxtail Millet Vacuum Cushion is very useful in skin cancer radiation treatment.
KEYWORD
skin cancer, bolus, skin dose, Foxtail Millet Vacuum Cushion