A high performance liquid chromatography using photoreduction fluorescence detection was described for the analysis of saikosaponins. Saikosaponins were separated on an NH2 column using acetonitrile and aqueous 2-tert-butylanthraquinone(t-BAQ) as mobile phase. Column effluent was passed through a 40cm PTFE capillary tube coiled around a 10W UV lamp to reduce t-BAQ to a highly fluorescent dihydroxyanthracene derivative which was detected by a fluorescence detector. The optimal concentration of t-BAQ was found to be 6X10-5M and the optimal reaction time to be 2 seconds. The detection limit for saikosaponin a and d by this method was found to be about 280ng and 80ng. The dynamic linear range was over two orders and the correlation coefficient of the calibration curve of them was 0.998.
|
|