KMID : 0603820170230020111
|
|
Journal of Experimental & Biomedical Science 2017 Volume.23 No. 2 p.111 ~ p.117
|
|
Comparison of Efficiency between Pre-storage and Post-storage filtration by Leukoreduction Blood Filter
|
|
Shin Geon-Sik
Kim Bo-Hee Kim Sung-Hoon Rhee Ki-Jong Kim Yoon-Suk
|
|
Abstract
|
|
|
Leukocytes in blood components are involved in diverse adverse transfusion reactions such as febrile non-hemolytic transfusion reactions. Therefore, leukocyte reduction is required to reduce these adverse reactions. The objective of this study was to compare the efficiency of pre-storage filtration and post-storage filtration. Filtration time, residual leukocyte count, RBC recovery, and hemolysis were assessed after pre-storage or post-storage filtration. Compared to pre-storage filtration, filtration time was prolonged and hemolysis was dramatically increased when post-storage filtration was performed. Residual leukocytes count and RBC recovery after post-storage filtration were similar with those obtained after pre-storage filtration. These results suggest that pre-storage filtration has better efficiency than post-storage filtration. These are thought to contribute to the production of better quality of leukoreduction blood components.
|
|
KEYWORD
|
|
Leukoreduction blood filter , Adverse transfusion reaction , Pre-storage filtration , Filtration efficiency
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|