Journal of the Korean Society for Therapeutic Radiology and Oncology 2008 Volume.26 No. 3 p.189 ~ p.194
|
|
Evaluation of a Water-based Bolus Device for Radiotherapy to the Extremities in Kaposi¡¯s Sarcoma Patients
|
|
Ahn Seung-Kwon
Lee Ik-Jae Suh Chang-Ok Kim Yong-Bae Cho Jae-Ho Kim Joo-Ho Cho Jeong-Hee Kim Dong-Wook Son Dong-Min Chang Young-Chea Song Tae-Soo
|
|
Abstract
|
|
|
Purpose: We designed a water-based bolus device for radiation therapy in Kaposi¡¯s sarcoma. This study
evaluated the usefulness of this new device and compared it with the currently used rice-based bolus.
Materials and Methods: We fashioned a polystyrene box and cut a hole in order to insert patient¡¯s extremities
while the patient was in the supine position. We used a vacuum-vinyl based polymer to reduce water leakage. Next, we eliminated air using a vacuum pump and a vacuum valve to reduce the air gap between the water and extremities in the vacuum-vinyl box. We performed CT scans to evaluate the density difference of the fabricated water-based bolus device when the device in which the rice-based bolus was placed directly, the rice-based bolus with polymer-vinyl packed rice, and the water were all put in. We analyzed the density change with the air gap volume using a planning system. In addition, we measured the homogeneity and dose in the low-extremities phantom, attached to six TLD, and wrapped film exposed in parallel-opposite fields with the LINAC under the same conditions as the set-up of the CT-simulator.
Results: The density value of the rice-based bolus with the rice put in directly was 14% lower than that of the
water-based bolus. Moreover, the value of the other experiments in the rice-based bolus with the polymer-vinyl packed rice showed an 18% reduction in density. The analysis of the EDR2 film revealed that the water-based bolus shows a more homogeneous dose plan, which was superior by 4.0¡4.4% to the rice-base bolus. The mean TLD readings of the rice-based bolus, with the rice put directly into the polystyrene box had a 3.4% higher density value. Moreover, the density value in the case of the rice-based bolus with polymer-vinyl packed rice had a 4.3% higher reading compared to the water-based bolus.
Conclusion: Our custom-made water-based bolus device increases the accuracy of the set-up by confirming
the treatment field. It also improves the accuracy of the therapy owing to the reduction of the air gap using a vacuum pump and a vacuum valve. This set-up represents a promising alternative device for delivering a homogenous dose to the target volume.
|
|
KEYWORD
|
|
Kaposi¡¯s sarcoma, Water-based bolus, Rice-based bolus
|
|