Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0948320060060020083
Konyang Medical Journal
2006 Volume.6 No. 2 p.83 ~ p.86
Evaluation of Corrected Dose with Inhomogeneous Tissue by Using CT Image
Kim Gha-Jung

Chung Weon-Kuu
Abstract
In radiation therapy, precise calculation of dose toward malignant tumors or normal tissue would be a critical factor in determining whether the treatment would be successful. The Radiation Treatment Planning (RTP) system is one of most effective methods to make it effective to the correction of dose due to CT number through converting linear attenuation coefficient to density of the inhomogeneous tissue by means of CT based reconstruction. In this study, we carried out the measurement of CT number and calculation of mass density by using RTP system and the homemade inhomogeneous tissue Phantom and the values were obtained with reference to water. Moreover, we intended to investigate the effectiveness and accuracy for the correction of inhomogeneous tissue by the CT number through comparing the measured dose (nC) and calculated dose (Percentage Depth Dose, PDD) used CT image during radiation exposure with RTP. The difference in mass density between the calculated tissue equivalent material and the true value was ranged from 0.005 g/cm3 to 0.069 g/cm3. A relative error between PDD of RTP and calculated dose obtained by radiation therapy of machine ranged from -2.8 to + 1.06% (effective range within 3%). In conclusion, we confirmed the effectiveness of correction for the inhomogeneous tissues through CT images. These results would be one of good information on the basic outline of Quality Assurance (QA) in RTP system.
KEYWORD
Correction of Inhomogenoeous Tissue, Radiation Treatment Planning System(RTPS), Hounsfield Unit(HU, CT number)
FullTexts / Linksout information
Listed journal information