Journal of Dental Rehabilitation and Applied Science
2015 Volume.31 No. 4 p.310 ~ p.315
Evaluation of fitness in implant screw as tightening torque in dental laboratory
¼Û¿µ±Õ:Song Young-Gyun
ÀúÀÚ¾øÀ½:No authors listed
Abstract
Purpose: The purpose of this study was to measure the tightening torque for dental implant in dental laboratory and to analyze of the effects of different tightening torque.

Materials and Methods: The tightening torque for dental implant in dental laboratory were measured by digital torque gauge. The length of abutment and analog were measured as tightening torque of manufacturer¡¯s instructions and the measured value. And the data were statically analyzed.

Results: The mean tightening torque of implant screw in dental laboratory was 1.563 ¡¾ 0.332 Ncm. The external type implant system of total length were showing no significant differences but the internal type implant system had difference significant (P < 0.05) when compared with tightening torque.

Conclusion: The implant prosthesis should be made under manufacturer¡¯s instructions especially as tightening torque of screw. For the fidelity of implant prosthesis, dental technician should learn how to use the torque gauge.
KEYWORD
implant torque controller, dental technician, implant-abutment, settlement