KMID : 1038920200190010001
|
|
Annals of Optometry and Contact Lens 2020 Volume.19 No. 1 p.1 ~ p.4
|
|
Comparison of Ocular Biometry Using Ultrasound and Swept-source Optical Coherence Tomography in Severe Cataract
|
|
Oh Hea-Young
Choi Jeong-Hoon Chung Young-Kwon
|
|
Abstract
|
|
|
Purpose: To compare the ocular biometry and prediction of postoperative refractive outcomes using IOL Master¢ç 700 (Carl Zeiss Meditec AG, Jena, Germany) and A-scan (Aviso¢ç, Quantel Medical, Cournon-d'Auvergne, France) in severe cataract.
Methods: Seventy nine eyes with severe cataract under visual acuity of 1.7 logMAR were scheduled for cataract surgery. Axial length and anterior chamber depth were measured preoperatively with IOL Master¢ç 700 and A-scan. Mean absolute error was obtained with the difference between preoperative predictive refraction and postoperative refraction outcome using autorefraction.
Results: Forty three cases received phacoemulsification and 36 received extracapsular cataract extraction. The mean absolute error between IOL Master¢ç 700 and A-scan in 20 eyes received phacoemulsification and 11 eyes received extracapsular cataract extraction (ECCE) was not significant (p < 0.001 and p < 0.001, respectively). IOL Master¢ç 700 measured mean axial lengths and anterior chamber depths longer than A-scan in phacoemulsification group (p < 0.001 and p < 0.001, respectively). However, there was no significant difference in mean axial length of ECCE group between two devices (p = 0.350).
Conclusions: There was no significant difference between IOL Master¢ç 700 and A-scan in prediction of refractive outcomes after surgery of severe cataract under visual acuity of 1.7 logMAR. If available, IOL Master¢ç 700 can be considered as reliable method for prediction of postoperative refractive outcomes in cases of severe cataract.
|
|
KEYWORD
|
|
A-scan, Cataract surgery, IOL Master¢ç 700, Refractive error, Visual acuity
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|