Journal of Agricultral Medicine & Community Health
2022 Volume.47 No. 4 p.220 ~ p.228
Survey of Point-of-Care Testing Devices for Pre-hospital EMS Equipment of Glucometer on Local Area
Kim Byeong-Hyun

Choi Dai-Hai
Ha Gyoung-Yim
Abstract
Objective: glucometer is one of POCT(point-of-care testing) devices used by EMS. Accurate measurement value is important for patient decision making. Quality control is necessary for this, but most EMS are still using glucometer without quality control it. Therefore, management status and quality control status of glucometers used by 119 paramedics waswere investigated.

Methods: 6 and 8 of glucometers and blood glucose test strips were submitted from 4 safety centers in 2010 and 8 safety centers in 2012, respectively. Glucometers submitted by local safety centers were analyzed in 2010 and 2012. The control glucometer, which was well managed at A University Hospital, was assumed to be accurate for at blood glucose measurement. Glucose values of various quality control materials control solutions were measured and analyzed using linear regression and Pearson¡¯s correlation test. Error rates were obtained via error tolerance range required by the 2011 ISO(International Organization for Standardization) guideline. For statistics, Excel and MedCalC programs were used.

Results: mong 14 safety centers in the region, 4 centers(28.5%) in 2010 and 8 centers(57.1%) in 2012 submitted 6 and 8 glucometers. In the first evaluation in 2010, 2 out of 6(33.3%) glucometers were untestable due to error, and 45 out of 6(66.683.3%) blood glucose test strips¡¯ validity were expired. In the second evaluation in 2012, 1 out of 8(13.5%) glucometers were untestable due to lack of battery, and 2 out of 8(25%) blood glucose test strips¡¯ validity were expired. All correlation coefficients were above 0.989 in 2010 and above 0.997 in 2012. Error rates were generally high when using expired blood glucose test strips.

Conclusion: Among 119 paramedics¡¯ guideline, quantity and operation status of POCT devices are well managed, but quality control is not. Therefore, it is necessary to add management status and quality control status of POCT devices to the EMS¡¯ field work guideline.
KEYWORD
Pre-hospital EMS, Glucometer, Quality assurance, POCT management