KMID : 1149020180200010033
Journal of Korean Society of Computed Tomographic Technology
2018 Volume.20 No. 1 p.33 ~ p.50
O-MAR, SBI image comparison & analysis evaluation about patients who have metal artifical implants in the knee joint CT scan
Kim Hae-Sol

Han Hyun-Jong
Kim Jae-Young
Park Cheon-Pil
Chang Sang-Gyu
Choi Jung-Won
Kim Jang-Seob
Abstract
Purpose : In the CT images, the Metal artifact causes disruption in imaging diagnosis. The study analyzes SBI Data images and O-MAR Data images of IQon CT (Philips) that reduces metal artifact, the study finds ideal Energy level that not have distortion of images for diagnosis among Energy levels of SBI Data.

Object and method : The study is targeted 20 peoples (The mean age - 46.3 years / The sex ratio - 14 males and 6 females) with the knee arthroplasty surgery of patellar fracture, ligament injury of knee, tibial fracture ? who examined the IQon CT scan from July 1, 2016 to January 31, 2017. The study acquired the data was applied O-MAR Reconstruction, SBI Reconstruction technique. The scan parameter is 120kVp, 120mAs, Pitch 0.391, Thickness 3mm, Rotation time 0.4 (sec).

Result : The SNR average value of the Knee joint Conventional data images is 7.79, the CNR average value is 19.09, HU average value is 312.93 SD average value is 45.86 the SNR average value of the O-MAR data images is 1.72, the CNR average value is 3.79 HU average value is 188.75 SD average value is 147.03 the SNR average value of the SBI Data 16.77, CNR average value is 22.65 HU average value is 76.22 SD average value is 15.84

Conclusion : In the study, the O-MAR image is effective to reduce the Metal artifact than conventional images. But, it was distortion of image reconstruction of soft tissue. On the other hand, the SBI data image is less distortion of image of Metal, soft tissue, bone to control Energy level.