KMID : 1164220200320010053
|
|
Journal of Korean Society for Radiotherapeutic Technology 2020 Volume.32 No. 1 p.53 ~ p.59
|
|
The Study on the Effect of Target Volume in DQA based on MLC log file
|
|
Shin Dong-Jin
Jung Dong-Min Cho Kang-Chul Kim Ji-Hoon Yoon Jong-Won Cho Jeong-Hee
|
|
Abstract
|
|
|
Purpose: The purpose of this study is to compare and analyze the difference between the MLC log file-based software (Mobius) and the conventional phantom-ionization chamber (ArcCheck) dose verification method according to the change of target volume.
Material and method: Radius 0.25cm, 0.5cm, 1cm, 2cm, 3cm, 4cm, 5cm, 6cm, 7cm, 8cm, 9cm, 10cm with a Sphere-shaped target Twelve plans were created and dose verification using Mobius and ArcCheck was conducted three times each. The irradiated data were compared and analyzed using the point dose error value and the gamma passing rate (3%/3mm) as evaluation indicators.
Result: Mobius point dose error values were -9.87% at a radius of 0.25cm and -4.39% at 0.5cm, and the error value was within 3% at the remaining target volume. The gamma passing rate was 95% at a radius of 9cm and 93.9% at 10cm, and a passing rate of more than 95% was shown in the remaining target volume. In ArcCheck, the average error value of the point dose was about 2% in all target volumes. The gamma passing rate also showed a pass rate of 98% or more in all target volumes.
Conclusion: For small targets with a radius of 0.5cm or less or a large target with a radius of 9cm or more, considering the uncertainty of DQA based on MLC log files, phantom-ionized DQA is used in complementary ways to include point dose, gamma index, DVH, and target coverage. It is believed that it is desirable to verify the dose delivery through a comprehensive analysis.
|
|
KEYWORD
|
|
target volume, Mobius, ArcCheck, DQA(Delivery Quality Assurance), point dose, gamma passing rate
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|