Journal of Korean Society for Radiotherapeutic Technology
2022 Volume.34 No. 1 p.31 ~ p.42
Evaluation of surface dose comparison by treatment equipment
Choi Eun-Ha

Yoon Bo-Reum
Park Byoung-Suk
An Ye-Chan
Park Myung-Hwan
Park Yong-Cheol
Abstract
Purpose: This study measures and compares the surface dose values in the virtual target volume using Tomotherapy, Halcyon, and TrueBeam equipment using 6MV-Flattening Filter-Free(FFF) energy.

Materials and Methods: CT scan was performed under three conditions of without bolus, 0.5 cm bolus, and 1 cm bolus using an IMRT phantom (IBA, Germany). The Planning Target Volume (PTV) was set at the virtual target depth, and the treatment plan was established at 200 cGy at a time. For surface dosimetry, the Gafchromic EBT3 film was placed in the same section as the treatment planning system and repeated measurements were performed 10 times and then analyzed.

Result: As a result of measuring the surface dose for each equipment, without, 0.5 cm, 1 cm bolus is in this order, and the result of Tomotherapy is 115.2¡¾2.0 cGy, 194.4¡¾3.3 cGy, 200.7¡¾2.9 cGy, The result in Halcyon was 104.7¡¾3.0 cGy, 180.1¡¾10.8 cGy, 187.0¡¾10.1 cGy, and the result in TrueBeam was 92.4¡¾3.2 cGy, 148.6¡¾5.7 cGy, 155.8¡¾6.1 cGy, In all three conditions, the same as the treatment planning system, Tomotherapy, Halcyon, TreuBeam was measured highly in that order.

Conclusion: Higher surface doses were measured in Tomotherapy and Halcyon compared to TrueBeam equipment. If the characteristics of each equipment are considered according to the treatment site and treatment purpose, it is expected that the treatment efficiency of the patient will increase as well as the treatment satisfaction of the patient.
KEYWORD
Surface Dose, Halcyon, TrueBeam, Tomotherapy, Gafchromic EBT3